Triangle Biomedical Sciences, Inc.
Solvent Recovery
S
ystem
Operation And Service
Manual
SR12
Version 1.2
Serial # ___________________
Vista de página 0
1 2 3 4 5 6 ... 15 16

Resumo do Conteúdo

Página 1 - Operation And Service

Triangle Biomedical Sciences, Inc. Solvent Recovery System Operation And Service Manual SR12 Version 1.2 Serial # ___________________

Página 2 - Page 2

Page 10 IMPORTANT: IF CHLORINATED SOLVENTS ARE TO BE RECOVERED, YOU MUST USE THE WORKING TEMPERATURES GIVEN IN THE ACCEPTABLE DISTILLATION SOLVEN

Página 3 - Page 3

Page 11 Shellsol D70-Exxol D80-Sepisol KD70 190-257 - Shellsol A Solvesso 100 Sepisol A 162-180 - Shellsol AB-Solvesso 150 182-215 - Shellsol H

Página 4 - Page 4

Page 12 NOTE: Because the oil expands when it is heated leakage will occur if the oil reservoir is over filled. Annual - After 2,000 hours of oper

Página 5 - Page 5

Page 13 MALFUNCTIONS, CAUSES AND REMEDIES Problem: Unit is heating up to set temps. Unit will recycle off water, alcohol and xylene into one c

Página 6 - Page 6

Page 14 become clogged with dirt TECHNICAL DATA Maximum Capacity 12 Liters (3.2 gal) 19 Liters (5.0 gal) Type of Distillation Simple Simpl

Página 7 - Page 7

Page 15 SPARE PARTS Code Description 12 liter 19 Liter - Cover Gasket, Ether √ - - Cover Gasket, Ether - √ 239004 Cover Gasket, Xylene √

Página 8 - Page 8

Page 16 GLOSSARY Auto-ignition Point - The temperature at which a solvent will self ignite (no stimulus required). Baffle - A metal duct used to di

Página 9 - Page 9

Page 2 TRIANGLE BIOMEDICAL SCIENCES, INC. 3014 CROASDAILE DRIVE • DURHAM, NORTH CAROLINA 27705 • 919.384.9393 • FAX 919.384.9595 •

Página 10 - Page 10

Page 3 GLOSSARY ...15 ...

Página 11 - Page 11

Page 4 FOREWORD This manual is designed to exhibit the information needed for installing, operating and troubleshooting the TRIANGLE BIOMEDICAL SCIE

Página 12 - Page 12

Page 5 GENERAL INFORMATION The TBS Solvent Recovery System is designed for recovering solvents used in Histopathology with a boiling point be

Página 13 - Page 13

Page 6 TBS Solvent Recovery Systems have been designed and manufactured to be ignition resistant if operated in a highly volatile environment (explos

Página 14 - Page 14

Page 7 OPERATING PROCEDURES Purpose - The TBS Solvent Recovery System was manufactured for the purpose of recovering solvents used within the labor

Página 15 - Page 15

Page 8 the Working Thermostat dial setting. The setting should be reduced to 150°C the same for Alcohol, and you must adjust the Fractional Thermos

Página 16 - GLOSSARY

Page 9 8. Set the timer for distillation. For a full distilling chamber it will take 5-6 hours for complete distillation to be accomplished. Th